Ex Parte MooreDownload PDFPatent Trial and Appeal BoardDec 17, 201212341427 (P.T.A.B. Dec. 17, 2012) Copy Citation UNITED STATES PATENT AND TRADEMARK OFFICE __________ BEFORE THE PATENT TRIAL AND APPEAL BOARD __________ Ex parte THOMAS G. MOORE JR. __________ Appeal 2012-002406 Application 12/341,427 Technology Center 3700 __________ Before LORA M. GREEN, STEPHEN WALSH, and JACQUELINE WRIGHT BONILLA, Administrative Patent Judges. WALSH, Administrative Patent Judge. DECISION ON APPEAL This is an appeal under 35 U.S.C. § 134(a) from the rejection of claims directed to a phacoemulsion system for use in ophthalmic surgery. The Patent Examiner rejected the claims for anticipation and obviousness. We have jurisdiction under 35 U.S.C. § 6(b). We affirm-in-part. Appeal 2012-002406 Application 12/341,427 2 STATEMENT OF THE CASE “Ophthalmic surgery often involves cutting away or emulsifying tissues that need to be removed from the eye, such as in cataract surgery.” (Spec. 1, [0003].) “Surgery is performed using a phacoemulsion handpiece.” (Id.) According to the Specification, “[a]ll commercially available phacoemulsification handpieces operate at a single fixed resonant frequency or a single fixed resonant frequency for each direction of travel.” (Id. at 9, [0020].) “For example a handpiece capable of both longitudinal and torsional vibration may have one resonant frequency for longitudinal vibration and a second resonant frequency for torsional vibration.” (Id.) Claims 1-10 are on appeal. Claim 1 is representative and reads as follows (emphasis added): 1. A phacoemulsification system for use in ophthalmic surgery, the phacoemulsification system comprising: a phacoemulsification handpiece having a horn coupled to a transducer that is configured to convert alternating current into mechanical oscillation of the horn, a phacoemulsification needle with an aspiration passage that is attached to the horn, whereby the phacoemulsification needle is made to vibrate by oscillation of the horn, to provide for mechanical cutting of tissue and inducing cavitation proximate a tip of the phacoemulsification needle, to thereby cause fragmentation of tissue at the surgical site; and a control system includes associated drive circuitry in connection with the transducer of the phacoemulsification handpiece device, the control system being configured to selectively adjust a non-resonant operating frequency of the transducer and vary an operating frequency of the phacoemulsification needle, to thereby increase or decrease a mechanical cutting performance and a cavitational-induced performance of the phacoemulsification needle. The Examiner rejected the claims as follows: I. claim 1 under 35 U.S.C. § 112, second paragraph, as indefinite; Appeal 2012-002406 Application 12/341,427 3 II. claims 1-4, 6, 7, 9, and 10 under 35 U.S.C. § 102(b) as anticipated by Koch;1 III. claim 5 under 35 U.S.C. § 103(a) as unpatentable over Koch and Alexander;2 IV. claim 8 under 35 U.S.C. § 102(b) as anticipated by or, in the alternative under 35 U.S. C. § 103(a) as obvious over, Koch. I The Appeal Brief lists the indefiniteness rejection among the rejections for review (App. Br. 3), but neither the Appeal Brief nor the Reply Brief discusses the merits of the rejection. We therefore summarily affirm it. See MANUAL OF PATENT EXAMINING PROCEDURE § 1205.02 (“If a ground of rejection stated by the examiner is not addressed in the appellant’s brief, that ground of rejection will be summarily sustained by the Board.”); see also, Ex parte Frye, 2010 WL 889747 *4 (Bd. Pat. App. & Int. 2010) (precedential) (“If an appellant fails to present arguments on a particular issue—or, more broadly, on a particular rejection—the Board will not, as a general matter, unilaterally review those uncontested aspects of the rejection”). II The Issue Claim 1’s phacoemulsification system comprises a handpiece and a control system. It is undisputed that Koch described a phacoemulsion 1 Hans-Reinhard Koch, US 6,716,219 B1, issued April 6, 2004. 2 Aaron James Alexander, US 6,283,974 B1, issued Sept. 4, 2001. Appeal 2012-002406 Application 12/341,427 4 system comprising a handpiece and a control system. The issue is whether Koch described a control system “configured to selectively adjust a non- resonant operating frequency of the transducer and vary an operating frequency of the phacoemulsification needle” and thereby increase or decrease needle performance. Findings of Fact 1. Koch disclosed: In parallel to these procedures, the phaco-machine function parameters are monitored continuously. These function parameters can be optimized by the operator, which is known as “tuning.” A head unit used to shatter the natural lens of the eye operates, for example, at a nominal frequency of 40 kHz. In order to determine the optimum frequency and amplitude, the phaco-head unit is first placed into a testing vessel and optimized. This value is stored in the control unit. During the operation, however, the optimum frequency changes when the phaco-head unit comes into contact with the natural lens, so that a certain improvement or re-tuning is required. This re-tuning is performed either by the operator or automatically via a control unit test program. (Koch, col. 8, ll. 22-34.) Principles of Law To anticipate a claim, a prior art reference must disclose every limitation of the claimed invention, either explicitly or inherently. Anticipation is an issue of fact, and the question whether a claim limitation is inherent in a prior art reference is a factual issue. Appeal 2012-002406 Application 12/341,427 5 In re Schreiber, 128 F.3d 1473, 1477 (Fed. Cir. 1997) (citations omitted). To anticipate, “the reference need not satisfy an ipsissimis verbis test.” In re Gleave, 560 F.3d 1331, 1334 (Fed. Cir. 2009). If a claim is subject to two interpretations and one interpretation would render the claim unpatentable over the prior art, the proper course of action is for the Examiner to enter both a § 112, ¶2, rejection and a rejection over prior art based on an interpretation of the claim which renders the prior art applicable. Ex parte Ionescu, 222 USPQ 537, 540 (Bd. Pat. App. & Int. 1984). Analysis Claim 1 Appellant’s control system is configured to selectively adjust a non-resonant operating frequency of the transducer and vary an operating frequency of the phacoemulsification needle, to thereby increase or decrease a mechanical cutting performance and a cavitational-induced performance of the phacoemulsification needle. (App. Br. 9, claim 1, emphasis added.) It is undisputed that Koch described a control system configured to tune or adjust an operating frequency. (FF 1.) Koch explained that if the nominal frequency were 40 kHz, for example, but re-tuning was required during operation, a control system could perform the re-tuning automatically. (FF 1.) That is, the control system could vary the operating frequency automatically from an undesired to a desired frequency. Koch did not distinguish between resonant and non-resonant frequencies. Koch did Appeal 2012-002406 Application 12/341,427 6 not specify that an undesired frequency was resonant or non-resonant, nor did Koch specify that an operating frequency must be a resonant frequency. Appellant contends that “Koch is simply an example of prior art systems that tune a handpiece having a nominal resonant frequency to a specific resonant frequency of the specific handpiece and to the load placed on that handpiece during surgery, nothing more.” (App. Br. 4.) More specifically, Appellant argues: All the frequency change teachings of Koch are with respect to obtaining an optimal (resonant) frequency; Koch is completely silent regarding the claimed control system adjusting a non-resonant transducer operating frequency to change the mechanical cutting and cavitation of the phaco needle. Koch is intent on moving from a non- resonant frequency to a resonant frequency, as opposed to the claimed invention of adjusting a non-resonant operating frequency. (Id.) Appellant also contends that “Koch does not have a control device that adjusts a non-resonant frequency so that a desired level of mechanical and cavitational performance is achieved.” (Id. at 5.) Appellant does not offer evidence to support their argument that Koch’s frequencies were only resonant frequencies. On this factual question, evidence, not argument, is what is needed. See In re Pearson, 494 F.2d 1399, 1405 (CCPA 1974) (“Attorney’s argument in a brief cannot take the place of evidence”); In re Geisler, 116 F.3d 1465, 1471 (Fed. Cir. 1997) (same). Because Koch did not limit its frequencies solely to resonant frequencies, and Appellants provide no other reason to give Koch such a narrow reading, we decline to do so. The Examiner responds that “Koch is fully capable of adjusting to a non-resonant frequency . . . . A user can also control the controller to adjust non-resonant frequencies.” (Ans. 9.) We find the Examiner’s explanation Appeal 2012-002406 Application 12/341,427 7 reasonable. Thus, if even the example “nominal” 40 kHz operating frequency in Koch was a resonant frequency for that device, Koch explains that if the device were in fact operating at some other frequency, which could be a non-resonant frequency, the controller could automatically adjust the non-resonant frequency to the desired operating frequency. That operation would meet the claim limitation “adjust a non-resonant operating frequency of the transducer.” The other function, “vary an operating frequency of the phacoemulsification needle,” is broader than the first because it is not limited to varying a non-resonant frequency. Koch’s device had that capability because, for example, if the undesired frequency were a resonant frequency, Koch’s controller would automatically vary it to the desired frequency, e.g., 40 kHz. That is, even if the operating frequency was an undesired resonant frequency, Koch described retuning, or varying, an operating frequency to the desired one. We therefore agree with the Examiner that Koch described what is now claimed. The rejection of claim 1 is affirmed. Claims 2, 6, 9, and 10 have not been argued separately and therefore fall with claim 1. 37 C.F.R. § 41.37(c)(1)(vii). Claims 3 and 4 Claim 3 further limits the system of claim 1 by reciting “wherein the control system is configured to change the operating frequency of the transducer after oscillation at the same operating frequency for more than a maximum duration of time.” The Examiner found Koch described this configuration because “the surgeon may determine a maximum length of time and manually change the Appeal 2012-002406 Application 12/341,427 8 frequency after that time has passed.” (Ans. 6, citing Koch, col. 8, ll. 33- 34.) Appellant contends that “[t]he cited Koch section simply teaches re- tuning the device after contacting a load, such as a lens; there is nothing in Koch that even hints at changing a nonresonant operating frequency of the transducer after a maximum amount of time.” (App. Br. 5.) We agree with Appellant. Koch’s control system may be configured to allow a surgeon to change the frequency, but there is no evidence the control system itself is configured to change the frequency after a maximum duration of time. The rejection of claim 3 is reversed. As claim 4 is dependent from claim 3, the rejection of claim 4 must also be reversed. Claims 7 and 8 Claim 7 further limits the system of claim 1 by reciting “and the control system is configured to selectively adjust the operating frequency of the transducer to establish a desired ratio of the mechanical disruption level relative to the cavitation-induced emulsification level.” The Examiner found Koch’s system had this feature because it was configured to allow the surgeon to adjust the frequency “as desired and needed.” (Ans. 6, citing Koch, col. 8.) Appellant disputes the Examiner’s finding “as the surgeon is only able to adjust the frequency to tune the device[,] not as desired and needed.” (App. Br. 6.) Further, Appellant contends that it does not matter if Koch’s operator is capable of adjusting the frequency as desired, “because claim 7 claims adjusting the operating frequency to establish a desired ratio, not simply adjusting the frequency.” Appeal 2012-002406 Application 12/341,427 9 Koch’s control system may be configured to allow a surgeon to change the frequency, but the rejection does not explain how Koch’s control system itself is configured to adjust the frequency to select a desired ratio of mechanical disruption level relative to the cavitation-induced emulsification level. The rejection of claim 7 is reversed. As claim 8 is dependent from claim 7, the rejection of claim 8 must also be reversed. III Regarding the obviousness rejection of claim 5, Appellant relies only on the argument that claim 1 is not anticipated by Koch. As the evidence supports the anticipation rejection of claim 1, we affirm the rejection of claim 5. IV Claim 8 is dependent from claim 7. The evidence is insufficient to show that Koch described a controller “configured to selectively adjust the operating frequency of the transducer to establish a desired ratio of the mechanical disruption level relative to the cavitation-induced emulsification level.” The obviousness rejection did not address that difference between Koch and claim 8. The rejection is therefore reversed. SUMMARY We affirm the rejection of claim 1 under 35 U.S.C. § 112, second paragraph, as indefinite. We affirm the rejection of claims 1, 2, 6, 9, and 10 under 35 U.S.C. § 102(b) as anticipated by Koch. Appeal 2012-002406 Application 12/341,427 10 We reverse the rejection of claims 3, 4, and 7 under 35 U.S.C. § 102(b) as anticipated by Koch. We affirm the rejection of claim 5 under 35 U.S.C. § 103(a) as unpatentable over Koch and Alexander. We reverse the rejection of claim 8 under 35 U.S.C. § 102(b) as anticipated by or, in the alternative under 35 U.S. C. § 103(a) as obvious over, Koch. No time period for taking any subsequent action in connection with this appeal may be extended under 37 C.F.R. § 1.136(a). AFFIRMED-IN-PART lp Copy with citationCopy as parenthetical citation