Ex Parte Christopherson et alDownload PDFPatent Trial and Appeal BoardJun 27, 201712992511 (P.T.A.B. Jun. 27, 2017) Copy Citation United States Patent and Trademark Office UNITED STATES DEPARTMENT OF COMMERCE United States Patent and Trademark Office Address: COMMISSIONER FOR PATENTS P.O.Box 1450 Alexandria, Virginia 22313-1450 www.uspto.gov APPLICATION NO. FILING DATE FIRST NAMED INVENTOR ATTORNEY DOCKET NO. CONFIRMATION NO. 12/992,511 03/07/2011 Mark Christopherson 1618.102.102 7866 25281 7590 06/29/2017 DICKE, BILLIG & CZAJA FIFTH STREET TOWERS 100 SOUTH FIFTH STREET, SUITE 2250 MINNEAPOLIS, MN 55402 EXAMINER CATINA, MICHAEL ANTHONY ART UNIT PAPER NUMBER 3735 NOTIFICATION DATE DELIVERY MODE 06/29/2017 ELECTRONIC Please find below and/or attached an Office communication concerning this application or proceeding. The time period for reply, if any, is set in the attached communication. Notice of the Office communication was sent electronically on above-indicated "Notification Date" to the following e-mail address(es): USPTO.PATENTS @dbclaw.com dmorris@dbclaw.com DBCLAW-Docket@dbclaw.com PTOL-90A (Rev. 04/07) UNITED STATES PATENT AND TRADEMARK OFFICE BEFORE THE PATENT TRIAL AND APPEAL BOARD Ex parte MARK CHRISTOPHERSON, TIMOTHY HERBERT, and QUANNI, Appeal 2016-001105 Application 12/992,5111 Technology Center 3700 Before TAWEN CHANG, DEVON ZASTROW NEWMAN, and DAVID COTTA, Administrative Patent Judges. NEWMAN, Administrative Patent Judge. DECISION ON APPEAL This appeal under 35 U.S.C. § 134 involves claims to a method of treating respiratory dysfunction, including sleep apnea. The Examiner entered final rejections for obviousness. We have jurisdiction under 35 U.S.C. § 6(b). We affirm. 1 Appellants identify the real party in interest as Inspire Medical Systems, Inc. App. Br. 3. Appeal 2016-001105 Application 12/992,511 STATEMENT OF THE CASE Background Sleep apnea generally refers to the cessation of breathing during sleep. One type of sleep apnea, referred to as obstructive sleep apnea (OSA), is characterized by repetitive pauses in breathing during sleep due to the obstruction and/or collapse of the upper airway, and is usually accompanied by a reduction in blood oxygenation saturation. One treatment for obstructive sleep apnea has included the delivery of electrical stimulation to the hypoglossal nerve, located in the neck region under the chin. Such stimulation therapy activates the upper airway muscles to maintain upper airway patency. Spec. 113^1. In addition to the electrode that delivers therapy to the hypoglossal nerve, “obstructive sleep apnea treatment may include the use of a sensor . . . implanted in a region that has pressure continuity with the intrapleural space,” which surrounds the lungs. {Id. 1 5, Fig. 2 (showing intrapleural space 90 and lung 80).) The Specification explains that “[cjhanges in intrapleural pressure provide a characteristic respiratory effort waveform, and therefore the sensor enables respiratory effort waveform information to be utilized to determine increased respiratory effort, which is then used to control delivery of therapy in response to determined increases in respiratory effort.” {Id.) The Specification notes that sensors may be attached to the trachea or esophagus, implanted in the suprasternal notch (the space between the trachea and esophagus), positioned intercostally (i.e., between the ribs,) or “secured in a position for sensing pressure at the posterior side of the manubrium” (the upper part of the sternum). {Id.) 2 Appeal 2016-001105 Application 12/992,511 The Specification discloses “an implantable stimulation system for stimulating and monitoring soft tissue in a patient, and more particularly, the invention relates to a method and apparatus for sensing respiratory pressure to control delivery of therapy to nerve tissue.” Id. 12. The Claims Claims 1, 2, 4, 10-20, 44, 45, and 59-63 are on appeal. App. Br. 1. Claims 1 and 59 are illustrative and read as follows: 1. A method of treating a respiratory dysfunction, the method comprising: positioning a pressure sensor of a lead, via an intercostal access point lateral to the sternum on one side of a patient’s body, into an implanted location in which the pressure sensor is coupled relative to respiratory pressures of a pleural space and in which a single side of the pressure sensor provides a single direction pressure sensing element, wherein the single side directly faces, and overlies, a portion of the lungs; maintaining the pressure sensor in the implanted location via securing a portion of the lead that is proximal to, and spaced apart from, the pressure sensor relative to the patient’s body without directly securing the pressure sensor at the location of the pressure sensor; sensing, via the pressure sensor, the respiratory pressure; and triggering electrical stimulation of an airway-patency- related nerve, from an implantable pulse generator coupled to the pressure sensor, based on at least one parameter of the sensed respiratory pressure. App. Br. 28 (Claims App.). 59. A method of treating obstructive sleep apnea, the method comprising: 3 Appeal 2016-001105 Application 12/992,511 locating and forming a subcutaneous, intercostal access point between a pair of immediately adjacent ribs on a right side of a patient’s body lateral to the sternum; and inserting and advancing a distal portion of a sensor lead, including a pressure sensor, through the intercostal access point to implant the distal portion on the right side of the patient’s body to couple the pressure sensor relative to respiratory pressures of a pleural space with a single side of the pressure sensor oriented to directly face toward a lung, wherein the single side includes a single-direction sensing element, and wherein the distal portion is generally parallel to the adjacent ribs and pointing toward a centerline of the body with the pressure sensor located at or lateral to a costochondral joint of the adjacent ribs to overly the portion of the lung; implanting a pulse generator on the right side of the patient’s body and coupling the pressure sensor relative to the pulse generator; sensing, via the pressure sensor, the respiratory pressure to provide a respiratory waveform; and triggering electrical stimulation of an airway-patency- related nerve, from the pulse generator, based on at least one parameter of the respiratory waveform. Id. at 34 (Claims App.). 4 Appeal 2016-001105 Application 12/992,511 The Issues The following rejections are before us to review: Claims 1, 2, 4, 10-20, 44, 45, and 59-63 are rejected under pre-AIA 35 U.S.C. § 103(a) as obvious over Christopherson2 and Lee.3 Final Act. 3. Claims 11—13, 44, 45, 59, and 60 are rejected under pre-AIA 35 U.S.C. § 103(a) as obvious over Christopherson, Lee, and Bolea.4 Id. at 11. ANALYSIS Rejection over Christopherson and Lee The issue with respect to this rejection is: Does the evidence of record support the Examiner’s conclusion that the claims are obvious over the cited references? The Examiner has rejected claims 1, 2, 4, 10-20, 44, 45, and 59-63 as obvious over Christopherson and Lee. The Examiner finds that Christopherson discloses a method of treating a respiratory dysfunction comprising: positioning a sensor portion of a lead, via an intercostal access point, into an implanted location in which the sensor portion is coupled relative to respiratory pressures of a pleural space . . . and in which a single side of the pressure sensor provides a single direction pressure sensing element, wherein the single side directly faces and overlies a portion of the lungs (. . . wherein the sensor can sense from both directions but either porthole only senses the pressure from the single side); maintaining the pressure sensor in the implanted location via securing a portion of the lead that is proximal to and spaced apart from, the pressure sensor relative to the patient’s body 2 US 6,132,384, issued Oct. 17, 2000 (“Christopherson”). 3 US 2005/0074741 Al, issued Apr. 7, 2005 (“Lee”). 4 US 2008/0103545 Al, published May 1, 2008 (“Bolea”). 5 Appeal 2016-001105 Application 12/992,511 without directly securing the pressure sensor at the location of the pressure sensor (.. . where the anchors are secured outside the rib cage. The sensor is not secured in that it can be placed elsewhere other than through the manubrium, like the intercostal space . . . and the anchors still fix the sensor lead). sensing, via the sensor portion at the implanted location, at least one of a temperature and the respiratory pressure . . . ; and . . . triggering the application of an electrical stimulation signal, from an implantable pulse generator coupled to the sensor portion, based on the substantial change in the amplitude of at least one of the sensed temperature and the sensed respiratory pressure. Final Act. 3^4. In other words, the Examiner appears to find that Christopherson teaches all limitations of claim 1, but notes that it does not teach locating and forming the intercostal access point between a pair of immediately adjacent ribs on one side of a patient’s body; and inserting and advancing a distal portion of the sensor lead, including the sensor portion, through the intercostal access point to orient the distal portion to be generally parallel to the adjacent ribs and pointing toward a centerline of the body. Id. at 4.5 The Examiner finds that Lee “teaches implanting the sensor in the correct orientation (. . . the sensor is between adjacent ribs, parallel in orientation and pointing inward).” Id. The Examiner concludes that it would have been obvious to combine the device of Christopherson with Lee as both are known positions for respiratory sensors thus the combination would have resulted in the predictable outcome of positioning 5 We note that the above-cited limitations are in fact limitations of claim 11 rather than claim 1. We address Appellants’ argument with respect to this point later in our opinion. 6 Appeal 2016-001105 Application 12/992,511 Christopherson’s sensors in the recited manner and the combination is no more than the predictable use of prior art elements for their established purpose. Furthermore Christopherson discloses various positions as long as they are near or along the sternum. Id. at 4—5. As stated in In re Oetiker, 977 F.2d 1443, 1445 (Fed. Cir. 1992): [T]he examiner bears the initial burden ... of presenting a prima facie case of unpatentability. . . . After evidence or argument is submitted by the applicant in response, patentability is determined on the totality of the record, by a preponderance of evidence with due consideration to persuasiveness of argument. We select claim 1 as representative of the claims subject to this ground of rejection. 37 C.F.R. § 41.37(c)(l)(iv). Unless otherwise noted, we adopt the Examiner’s findings of fact, reasoning on scope and content of the prior art, and conclusions set out in the Final Action and Answer. We conclude that the Examiner has established a prima facie case that the method of claim 1 would have been obvious over Christopherson and Lee. Appellants have not produced evidence showing, or persuasively argued, that the Examiner’s determinations are incorrect. Only those arguments made by Appellants in the Briefs have been considered in this Decision. Arguments not presented in the Briefs are waived. See 37 C.F.R. § 41.37(c)(l)(iv) (2015). Appellants argue Christopherson does not provide sufficient information or illustration regarding an intercostal positioning to guide the skilled artisan to use this positioning with a pressure sensor as claimed. App. Br. 9. Appellants further argue that the skilled artisan would not be motivated to combine the teachings of Lee with those of Christopherson, 7 Appeal 2016-001105 Application 12/992,511 because “the placement of the pressure sensor in Christopherson is for a [different] purpose (measuring pressure) than the purpose (measuring impedance) of the placement of the electrode assembly 308 in Lee.” Id. at 16. We are not persuaded. Christopherson teaches that its “pressure sensor 60 is surgically implanted in a region that “has pressure continuity with the intrapleural space such as ... an intercostal placement” and that the location for placement of the sensor is at least in part, chosen as a function of delay, i.e. propagation time of a pressure waveform characteristic of respiratory effort propagating from the respiratory point of origin to the sensor position and as a function of the amount of filtering necessary to achieve a usable sensed signal at a particular location, i.e. filtering necessary to remove waveforms other than the waveform of the sensed characteristic, such as cardiac waveform activity. Christopherson 8:54—9:7; see also id. at 2:28—33 (prior art teaches intercostal placement of sensor). Christopherson further teaches that preferable placement of the sensor is implanted through a drilled hole in the manubrium 63 as shown in FIGS. 10a and 10b. However, the sensor assembly 115 could be implanted through the sternum 64 at any location thereof or through any other bone such that the sensing element 126 is in communication with the intrathoracic region or a region with pressure changes characteristic of respiratory effort. Id. at 11:42-49. Accordingly, because Christopherson recognizes both intercostal positioning and positioning along or through the sternum as positions providing sufficient continuity with the pleural space as to permit proper calculation of the waveform, and expressly instructs the skilled artisan that these positions may be used, we are persuaded that the artisan 8 Appeal 2016-001105 Application 12/992,511 would be able to achieve the claimed position (i.e., “via an intercostal access point lateral to the sternum on one side of a patient’s body”) using the guidance of Christopherson alone.6 We note, however, that the Examiner relies upon Lee to teach positioning of the sensor. Ans. 2. We agree with the Examiner that Lee also teaches intercostal positioning (see Lee, Lig. 3,1 56) of a sensing and/or stimulation device and further find that Christopherson’s teaching is not limited to pressure sensors. Christopherson 2:59—60 (pressure sensing element is one embodiment), 4:39-49 (invention broadly described as a system including an implantable sensing element for sensing “a characteristic of respiratory effort”). Christopherson specifically teaches that the characteristic monitored by Lee, intrathoracic impedance, can also be used to obtain respiratory effort waveforms that are sufficiently identifiable for use with its pressure sensors to monitor respiratory effort: “[rjespiratory effort waveforms having similar identifiable characteristics can be provided by monitoring other physiological signals such as intrathoracic pressure, intrathoracic impedance or electromyographic potentials. ” Id. at 7:12—16. Lor this reason we are persuaded that the skilled artisan would have reason to combine the teachings of Christopherson and Lee and agree with the Examiner that there “are limited positions in which a sensor can be placed in the thoracic cavity and, as such, it would not require undue experimentation to determine an optimum position for lung 6 The Board may rely upon fewer than all the references cited by the Examiner. See In re May, 574 L.2d 1082, 1090 (CCPA 1978). 9 Appeal 2016-001105 Application 12/992,511 monitoring especially given that intercostal positioning of sensors is known as shown in Lee.” Ans. 2.7 Appellants next argue that Christopherson’s conflicts with the claim requirement that “a single side of the pressure sensor provides a single direction pressure sensing element, wherein the single side directly faces and overlies a portion of the lungs” because neither of the disclosed directions of the porthole pressure sensor directly faces the lungs but instead “face[s] generally parallel to the lungs.” App. Br. 9—10. Appellants argue, the skilled artisan would understand that “the sensing section 120 in Christopherson detects pressure from two opposite directions via the oppositely oriented portholes 128.” Id at 10. We are not persuaded. Claim 1 requires that “a single side of the pressure sensor provides a single-direction pressure sensing element, wherein the single side directly faces, and overlies, a portion of the lungs.” Figures 11a and 1 lb of Christopherson are reproduced below: 7 We acknowledge Appellants’ argument that claim 11, not claim 1, contains the language that the Examiner finds is taught by Lee and that Appellants argue this fact renders the Examiner’s proposed combination of Christopherson and Lee without “rationale.” App. Br. 15. We are not persuaded as Lee is analogous art that provides design options for to the skilled artisan to consult as how to place a sensor with regard to orientation and placement in the intercostal space. See In re Kuhle, 526 F.2d 553, 555 (CCPA 1975) (“[T]he particular placement of the contact provides no novel or unexpected result.”). Furthermore, claim 11 depends from claim 1. To the extent Christopherson and Lee render claim 11 obvious, they also render the broader claim 1 obvious. 10 Appeal 2016-001105 Application 12/992,511 \ FIG.IIa US FIG.IIb Figures 11a and 1 lb of Christopherson depict, respectively, a side view and a cutaway view of an embodiment of its pressure sensor. Christopherson 5:33—37. Referring to Figure 11, Christopherson discloses that the “pressure sensor assembly 115 .. . includes a sensing section 120.” Id. 9:12—13. Christopherson teaches that “[t]he relative pressure sensing element 126 senses respirational pressures through the use of piezo-electric crystals attached to a sensor diaphragm lying perpendicular to a longitudinal axis 125 extending through assembly 115.” Id. at 9:18—22 (emphasis added). Christopherson’s Figure 10a is reproduced below: Christopherson’s Figure 10a is an “illustration^ showing [a] position[] or configuration^ for mounting the sensor.” Id. at 5:29—32. As shown in Figure 10a, a sensor diaphragm lying perpendicular to 62 /, 115 FIG.10a 11 Appeal 2016-001105 Application 12/992,511 the longitudinal axis of assembly 115 appears to face the direction of the lung. We further agree with the Examiner that claim 1 does not require that the pressure sensor receive input only from the side it faces, but simply that it senses pressure at the location of the sensor. Final Act. 14; Ans. 3. As demonstrated above in Figure 10a of Christoperson, the sensor faces the patient’s lung sufficiently to detect pressure. Ans. 3. As the Examiner further points out, Fee teaches a sensor in a disordered breathing classification and/or discrimination system overlying a portion of the lung. Ans. 3; Fee Fig. 3. Appellants argue that a skilled artisan viewing Fee’s Figure 3 would “likely observe that the placement of the electrode assembly 307 is not likely to significantly overlie lung tissue.” Appeal Br. 14 (emphasis added). However, we are not persuaded because we agree with the Examiner that claim 1 does not specify a “degree to which the sensor must overlie the lung.” Ans. 3. In sum, we find the Examiner has established a prima facie case of obviousness with respect to the limitation and that the burden shifts to Appellants to prove otherwise. In re Oetiker, 977 F.2d at 1445. Appellants have provided only unpersuasive attorney argument. See In re Geisler, 116 F.3d 1465, 1470 (Fed. Cir. 1997) (“[Attorney argument [is] not the kind of factual evidence that is required to rebut a prima facie case of obviousness”). Appellants next argue that certain embodiments of Christopherson do not teach the claim limitation: “maintaining the pressure sensor in the implanted location via securing a portion of the lead that is proximal to, and 12 Appeal 2016-001105 Application 12/992,511 spaced apart from, the pressure sensor relative to the patient’s body without directly securing the pressure sensor at the location of the pressure sensor.” App. Br. 11—13. This argument is not persuasive because, as the Examiner’s notes, Christopherson discloses at least one embodiment comprising anchors on the lead of the sensor that secure the lead and maintain the sensor’s location but do not directly secure the sensor body itself. Ans. 3; Christopherson Figs. 10c (anchoring elements 178 securing a portion of the lead that is proximal to but spaced apart from the sensing element 197) and 11a (pressure sensor assembly or sensor lead comprising a sensing section 120 and a separate “lead anchoring” section 122). This is sufficient to establish obviousness, as “[i]t is well settled that a prior art reference is relevant for all that it teaches to those of ordinary skill in the art.” In re Fritch, 972 F.2d 1260, 1264 (Fed. Cir. 1992). Appellants next argue that Fee fails to cure the deficiencies of Christopherson such that the combination of Fee and Christopherson does not render claim 1 obvious. App. Br. 13—15. We are not persuaded for the reasons discussed above. Finally, we note but are not persuaded by Appellants’ argument in the Reply Brief that the Examiner’s obviousness rejection is based on a combination of different embodiments in Christopherson and Fee. (Reply 2, 3—4.) There is no requirement in an obviousness rejection that all of the limitations be disclosed in a single embodiment. Moreover, while Appellants argue that the various embodiments conflict with or are inconsistent with each other, Appellants have provided no evidence or persuasive argument that any such differences in the embodiments would render their combination inoperable or non- obvious. Cf. Orthopedic Equip. Co. v. United States, 702 F.2d 1005, 1013 13 Appeal 2016-001105 Application 12/992,511 (Fed. Cir. 1983) (explaining that the fact that two prior art devices could not physically be combined does not mean that those skilled in the art would have found combining features of the two devices to be nonobvious). Accordingly, we affirm the rejection of claim 1. Claim 11 Appellants argue that Lee does not teach ‘“locating and forming an intercostal access point’ at the location ‘between a pair of immediately adjacent ribs’” because it is relatively silent “regarding a manner and/or location of securing electrode assembly 307.” App. Br. 17. This argument is not persuasive because Figure 3 of Lee shows the sensor in the precise position recited in the claim, and paragraph 55 describes the device of Figure 3 as follows (emphasis added): The ITCS device may, for example, be implanted subcutaneously such that all or selected elements of the device are positioned on the patient’s front, back, side, or other body locations suitable for sensing cardiac activity and delivering cardiac stimulation therapy. Elements of the ITCS device may be located at several different body locations, such as in the chest, abdominal, or subclavian region with electrode elements respectively positioned at different regions near, around, in, or on the heart. We find that with this description and the teachings of Lee and Christopherson regarding intercostal placement discussed above, the intercostal implantation as recited of claim 11 would have been obvious to the skilled artisan. Appellants further argue that, in contrast to the Examiner’s assertion that Appellants have not disclosed an advantage to placing the device on the right side of the body, paragraphs 67—69 of the Specification demonstrate the advantage of placing the sensor on the right side of the body, which is to 14 Appeal 2016-001105 Application 12/992,511 “maximize its spacing from the cardiac activity associated with the apex of the heart and with the left ventricle, which are on the left side of the body. . . . [thereby minimizing] the impact of cardiac artifacts on sensing respiratory pressures via the pressure sensor.” App Br. 17; Spec. ^fl[ 67—69. This argument is not persuasive. Christopherson explicitly teaches that “the location for placement of the sensor is, at least in part, chosen . . . as a function of the amount of filtering necessary ... to remove waveforms other than the waveform of the sensed characteristic, such as cardiac waveform activity.'” Christopherson 8:66—9:7 (emphasis added). Accordingly, it would have been obvious for the skilled artisan “to optimize the location of the sensor to get less interference as there are limited positions with[in] the chest cavity to try” (Ans. 4) and the prior art has indicated that any of them are suitable so long as the sensor is in a position that “has pressure continuity with the intrapleural space” and a stable waveform signal can be obtained. Christopherson 8:54—9:7. For these reasons we are also unpersuaded by Appellants’ argument that the Examiner’s citation to Figure 3 and paragraphs 63—64 of Lee as teaching implantation of the sensor in the claimed orientation is hindsight reconstruction. Indeed, [a]ny judgment on obviousness is in a sense necessarily a reconstruction based upon hindsight reasoning, but so long as it takes into account only knowledge which was within the level of ordinary skill at the time the claimed invention was made and does not include knowledge gleaned only from applicant’s disclosure, such a reconstruction is proper. In re McLaughlin, 443 F.2d 1392, 1395, (CCPA 1971). The Examiner’s rejection is properly based on prior art. 15 Appeal 2016-001105 Application 12/992,511 To the extent Appellants argue that the placement of claim 11 permits them to obtain unexpected results such that the invention is not obvious, Appellants’ argument is further unpersuasive without evidence of results superior to the closest prior art, which was not provided. See In re Baxter TravenolLabs., 952 F.2d 388, 392 (Fed. Cir. 1991) (“[W]hen unexpected results are used as evidence of nonobviousness, the results must be shown to be unexpected compared with the closest prior art”). Neither have Appellants shown that any superior results are unexpected, given Christopherson’s teaching that minimizing cardiac waveform activity is desirable in determining sensor placement. Appellants’ attorney argument alone is unpersuasive. See In re Geisler, 116 F.3d at 1470 (“[Attorney argument [is] not the kind of factual evidence that is required to rebut a prima facie case of obviousness”). Accordingly, we affirm the rejection of claim 11. Claims 12, 15, and 20 Appellants argue that the location of the sensor placement in claim 12 — at or lateral to the costochondral joint — is not obvious based on the Examiner’s combination of prior art references. App. Br. 19. According to Appellants, the positioning of the disordered breathing classification system in Lee’s Figure 3 is “such that electrode assembly 307 is apparently located medially of the costochondral joint of the ribs (and their associated costal cartilages) immediately adjacent the electrode assembly 307” as opposed to the claimed ‘“at or lateral to the costochondral joint of the adjacent ribs.’” Id. Similarly, Appellants argue that the location of the sensor in claim 15, “between the fifth and sixth rib” is nonobvious because it “helps to 16 Appeal 2016-001105 Application 12/992,511 overcome the stated problem of avoiding cardiac artifacts . . . and/or avoiding a more invasive pectoral location for sensor implant” thereby “significantly contributing] to alleviating respiratory dysfunction.” Id. at 20. Appellants additionally argue that the placement of the sensor and anchor lead in claim 20, in which the “fixed first anchor is spaced apart from the pressure sensor about 3 to about 9 centimeters along a length of the sensor lead” is not obvious because the Specification “articulate[s] an advantage and purpose . . . that provides a meaningful difference” provided by this arrangement and neither cited reference discloses it. Id. at 21—22. Once again we are unpersuaded due to the disclosures of Christopherson and Lee, as discussed above, that a sensor as claimed could be placed in multiple locations near the sternum based on consideration of factors including the minimization of cardiac artifacts, as well as the ability of the skilled artisan to optimize placement based on the measurement to be obtained. To the extent that Appellants argue that their claimed placement provides an unexpected advantage over the prior art, this argument alone is unpersuasive without evidence that this particular placement provides an unexpected advantage as compared to the closest prior art, and that such a benefit was indeed “unexpected.” In re Geisler, 116 F.3d at 1470. On this record, the evidence supports the Examiner’s finding that the combination of Christopherson (e.g., Figure 10b—lOd) and Lee suggests: “arranging the implanted location of the pressure sensor to be at or lateral to costochondral joint of the adjacent ribs,” “wherein the adjacent pair of ribs comprises the fifth uppermost rib and the sixth uppermost rib” and wherein a 17 Appeal 2016-001105 Application 12/992,511 “fixed first anchor is spaced apart from the pressure sensor about 3 to about 9 centimeters along a length of the sensor lead.” Accordingly, we affirm the rejection of claims 12, 15, and 20. Claim 17 Appellants argue that the claim limitation “securing, via a fixed first anchor on the sensor lead that is proximal to the sensor portion, the sensor lead relative to a subcutaneous connective tissue that is external to the intercostal access point” is not disclosed by Christopherson. App. Br. 21. The Examiner cited Figure 10 as teaching that the fixed first anchor of the sensor lead is proximal to the sensor portion and the lead is external to the intercostal access point and that the “anchors are secured outside the rib cage.” Final Act. 4. Figure 10 is an embodiment reflecting a configuration with an anchor outside the rib cage and alongside the right arm: 185 Appellants provide no evidence or persuasive explanation why Figure 10b does not teach the recited configuration or why the proposed combination does not render claim 17 obvious. Presenting a statement that merely points out what a claim recites is not an argument for the separate patentability of 18 Appeal 2016-001105 Application 12/992,511 the claims. 37 C.F.R. § 41.37(c)(l)(iv). See also In reLovin, 652 F.3d 1349, 1357 (Fed. Cir. 2011) (noting that separately arguing a claim requires “more substantive arguments in an appeal brief than a mere recitation of the claim elements and a naked assertion that the corresponding elements were not found in the prior art.”). Accordingly we affirm the rejection of claim 17. Claims 59, 60, and 61 Appellants rely on their prior arguments in support of their request to reverse the rejection of these claims and provide no separate arguments. In re Lovin, 652 F.3d at 1357. Accordingly, we affirm their rejection. Conclusion of Law We affirm the rejection of claims 1, 11, 12, 15, 17, 20, and 59-61 as obvious over Christopherson and Lee. Claims 2, 4, 10, 13, 14, 16, 18, 19, 44, 45, 62, and 63 were not argued separately; their rejection is also affirmed. 37 C.F.R. § 41.37(c)(l)(iv). Rejection over Christopherson, Lee, and Bolea The Examiner has rejected claims 11—13, 44, 45, 59, and 60 as obvious over Christopherson, Lee, and Bolea. The Examiner finds that: Christopherson as modified does not specifically disclose placing the lead and IPG on the right side of the patient. Bolea, however, teaches implanting the IPG on the right side and the respiration sensor between the 5th and 6th rib ... . Therefore it would have been obvious ... to combine the device of Christopherson with the teachings of Bolea in order to leave space for a pacemaker traditionally implanted on the left side. Final Act. 11. 19 Appeal 2016-001105 Application 12/992,511 We select claim 11 as representative of the claims subject to this ground of rejection. We adopt the Examiner’s findings of fact, reasoning on scope and content of the prior art, and conclusions set out in the Final Action and Answer. We conclude that the Examiner has established a prima facie case that the method of claim 11 would have been obvious over Christopherson, Lee, and Bolea. Appellants have not produced evidence showing, or persuasively argued, that the Examiner’s determinations are incorrect. Appellants reiterate their prior arguments regarding the teachings of Christopherson and Lee. For the reasons above, those arguments are unpersuasive. Appellants also argue that Bolea does not cure the deficiencies of Christopherson and Lee, because it fails to disclose any arrangement corresponding to [jlocating and forming the intercostal access point between a pair of immediately adjacent ribs on a right side of the patient’s body; and inserting and advancing a distal portion of the sensor lead, including the pressure sensor, through the intercostal access point to orient the distal portion on the right side of the patient’s body to be generally parallel to the adjacent ribs and pointing toward a centerline of the body. App. Br. 24. According to Appellants, Bolea’s “mention of the ‘right side of the patient’s body’” is insufficient to teach this arrangement. Id. We are not persuaded. Bolea explicitly teaches placement of its implantable neurostimulator (INS) sensor on the right side of the patient: the bio-Z vector may be arranged such that the proximal combination electrode is located just to the right of and above the xiphoid below the pectoral muscle between the 5th and 6th ribs and the distal current-carrying electrode is located midlateral between the 7th and 8th ribs, with the distal voltage sensing electrode positioned between the 6th and 7th ribs 10 mm proximal of the distal current-carrying electrode. This 20 Appeal 2016-001105 Application 12/992,511 arrangement places the electrodes along the interface between the right lung, diaphragm and liver on the right side of the thoracic cavity. Bolea 1176; see also id. 177—179. In addition, Christopherson and Lee both teach placement in multiple locations adjacent to the sternum, as discussed above. As further discussed above, to the extent Appellants allege unexpected superior effect of the claimed placement, this is unsupported attorney argument. Accordingly, we affirm the rejection. Claim 12 Appellants argue that because Bolea discloses a “trans-thoracic impedance measurement technique,” the skilled artisan would recognize that the proximal combination electrode is located nearly immediately adjacent the sternum and, therefore, cannot be “at or lateral to a costochondral joint of the adjacent ribs.” App. Br. 25. Appellants also argue “to the extent that the Final Office Action focuses on a single electrode as being between the fifth and sixth ribs, this particular electrode (the ‘proximal combination electrode’) is placed closer toward the Left side of the body, and therefore closer toward the likely position of the heart”. Id. We are not persuaded. As discussed above, Christopherson and Lee suggest a wide flexibility of placement. As noted by the Examiner, Bolea teaches positioning of the IPG on the right side and a respiration sensor between the 5th and 6th rib. Final Act. 11. Claim 12 recites “wherein positioning the pressure sensor comprises: arranging the implanted location of the pressure sensor to be at or lateral to a costochondral joint of the adjacent ribs”. App. Br. 29 (Claims App.). The positioning shown in Bolea teaches “at or lateral to a costochondral joint of the adjacent ribs” and is shown on the right side of the patient. Bolea Tflf 176—179, Fig. 37. 21 Appeal 2016-001105 Application 12/992,511 Appellants have not provided arguments or evidence that, when considered together with the Examiner’s reasonable fact-based conclusions, render the claim non-obvious. Accordingly, we affirm the rejection of claim 12. Appellants raise no further substantive argument regarding the rejected claims. Accordingly, we affirm the rejection of claims 13, 44, 45, 59, and 60. SUMMARY We affirm the rejection of claims 1, 2, 4, 10—20, 44, 45, and 59-63 under pre-AIA 35 U.S.C. § 103(a) as obvious over Christopherson and Lee. We affirm the rejection of claims 11—13, 44, 45, 59, and 60 under pre- AIA 35 U.S.C. § 103(a) as being obvious over Christopherson, Lee, and Bolea. No time period for taking any subsequent action in connection with this appeal may be extended under 37 C.L.R. § 1.136(a)(1). See 37 C.L.R. § 1.136(a)(l)(iv). ALLIRMED 22 Copy with citationCopy as parenthetical citation