Table 3-Performance Test Requirements
Work-rate test sequence | Duration per cycle (in minutes) | VO2 (L/min) | VCO2 (L/min) | Ve (L/min) | RF (breaths/min) |
1. Peak | 5 | 3.00 | 3.20 | 65.0 | 25 |
2. High | 15 | 2.00 | 1.80 | 44.0 | 20 |
3. Low | 10 | 0.50 | 0.40 | 20.0 | 12 |
VO2 = volume of oxygen consumed per minute; VCO2 = volume of carbon dioxide produced per minute.
Ve = ventilation rate in liters of air per minute; RF = respiratory frequency.
42 C.F.R. §84.305