Ex Parte Armoundas et alDownload PDFPatent Trial and Appeal BoardAug 8, 201713509390 (P.T.A.B. Aug. 8, 2017) Copy Citation United States Patent and Trademark Office UNITED STATES DEPARTMENT OF COMMERCE United States Patent and Trademark Office Address: COMMISSIONER FOR PATENTS P.O.Box 1450 Alexandria, Virginia 22313-1450 www.uspto.gov APPLICATION NO. FILING DATE FIRST NAMED INVENTOR ATTORNEY DOCKET NO. CONFIRMATION NO. 13/509,390 08/29/2012 Antonis Armoundas 125141.00489 1900 26710 7590 QUARLES & BRADY LLP Attn: IP Docket 411 E. WISCONSIN AVENUE SUITE 2350 MILWAUKEE, WI 53202-4426 EXAMINER PATTON, AMANDA K ART UNIT PAPER NUMBER 3762 NOTIFICATION DATE DELIVERY MODE 08/10/2017 ELECTRONIC Please find below and/or attached an Office communication concerning this application or proceeding. The time period for reply, if any, is set in the attached communication. Notice of the Office communication was sent electronically on above-indicated "Notification Date" to the following e-mail address(es): pat-dept@quarles.com PTOL-90A (Rev. 04/07) UNITED STATES PATENT AND TRADEMARK OFFICE BEFORE THE PATENT TRIAL AND APPEAL BOARD Ex parte ANTONIS ARMOUNDAS and ERIC WEISS Appeal 2016-006425 Application 13/509,390 Technology Center 3700 Before ERIC B. GRIMES, JEFFREY N. FREDMAN, and DEVON ZASTROW NEWMAN, Administrative Patent Judges. FREDMAN, Administrative Patent Judge. DECISION ON APPEAL This is an appeal1 under 35U.S.C. § 134 involving claims to a method for preemptively suppressing a heart rhythm disturbance in a patient’s heart. The Examiner rejected the claims as obvious. We have jurisdiction under 35 U.S.C. § 6(b). We reverse. Statement of the Case Background “Arrhythmias such as ventricular tachycardia and fibrillation are often caused by an electrical mechanism called reentry ... A major factor leading to the genesis of ventricular fibrillation during ischemia is dispersion of refractoriness” (Spec. | 6). “An important mechanism for enhancing 1 Appellants do not identify the Real Party in Interest. Appeal 2016-006425 Application 13/509,390 dispersion of refractory period is alternation of the action potential from beat to beat” (id. 17). “Action potential altemans involves an alternating sequence in which the shape of the action potential” “creates a situation in which a region of the myocardium has a long refractory period on an every other beat basis” (id. 1 8). “Thus, action potential altemans, which generally occurs in diseased tissue, can promote the development of reentry” (id.). “A reentrant waveform can be terminated by electrical pacing that is initiated within a specific period during reentrant excitation. This process gives rise to a new wave whose front collides with and annihilates the reentrant wave” (id. 115). “Thus, it would be highly desirable to be able to prevent arrhythmias from starting rather than terminating them after their initiation by administration of an electrical shock” (id. 116). The Claims Claims 1—5 and 7—21 are on appeal. Claim 1 is representative and reads as follows: 1. A method for preemptively suppressing a heart rhythm disturbance in a patient’s heart with an implanted device configured to generate electrical impulses, the steps of the method comprising: a) detecting cardiac electrical signals from the patient’s heart; b) measuring a beat-to-beat variability in the detected cardiac electrical signals and estimating at least one repolarization altemans parameter therefrom; c) determining a severity and likelihood of an imminent heart rhythm disturbance occurrence by comparing the at least one repolarization altemans parameter to a dynamically determined threshold value that is adaptively changed based on a noise level in the detected cardiac electrical signals, wherein 2 Appeal 2016-006425 Application 13/509,390 the severity and likelihood of an imminent heart rhythm disturbance occurrence is characterized by a dynamically changing level of repolarization altemans in the detected cardiac electrical signals; d) calibrating an electrical therapy plan using the determined severity and likelihood of the imminent heart rhythm disturbance occurrence; and e) delivering appropriately calibrated electrical impulses to the patient’s heart with the implanted device using the calibrated electrical therapy plan in order to preemptively suppress the severity and likelihood of the imminent heart rhythm disturbance occurrence. The Issues A. The Examiner rejected claims 1—5 and 10-20 under 35 U.S.C. § 103(a) as obvious over Armoundas,2 Farazi,3 Christini,4 and Krishnamachari5 (Ans. 2—5). B. The Examiner rejected claims 7 and 21 under 35 U.S.C. § 103(a) as obvious over Christini and Zhu6 (Ans. 5—7). C. The Examiner rejected claims 8 and 9 under 35 U.S.C. § 103(a) as obvious over Christini, Zhu, and Sharma7 (Ans. 8—9). 2 Armoundas et al., US 2007/0191890 Al, published Aug. 16, 2007. 3 Farazi, US 7,756,571 Bl, issued July 13, 2010. 4 Christini et al., US 6,915,156, issued July 5, 2005. 5 Krishnamachari, US 6,453,191 B2, issued Sept. 17, 2002. 6 Zhu et al., US 7,245,970 B2, issued July 17, 2007. 7 Sharma et al., US 2011/0105929 Al, published May 5, 2011. 3 Appeal 2016-006425 Application 13/509,390 A. 35 U.S.C. § 103(a) over Armoundas, Farazi, Christini, and Krishnamachari The Examiner finds: Armoundas discloses a method for preemptively suppressing a heart rhythm disturbance in a patient’s heart with an implanted device configured to generate electrical impulses, the steps of the method comprising: a) detecting cardiac electrical signals from the patient’s heart; b) measuring a beat-to-beat variability in the detected cardiac electrical signals; c) determining a severity and likelihood of an imminent heart rhythm disturbance occurrence by comparing the measured beat-to-beat variability to a threshold value; d) calibrating an electrical therapy plan using the determined severity and likelihood of the imminent heart rhythm disturbance occurrence; and e) delivering appropriately calibrated electrical impulses to the patient’s heart with the implanted device using the calibrated electrical therapy plan in order to preemptively suppress the severity and likelihood of the imminent heart rhythm disturbance occurrence. (Ans. 2—3). The Examiner finds “Farazi discloses a degree of altemans utilized as an index for impending ventricular arrhythmia and a dynamic threshold for said index . . . Christini discloses dynamic thresholds in general” {id. at 3). The Examiner finds it obvious to “modify Armoundas to include an alteman index and dynamic thresholding associated with t[he] index, as taught by Christini and Farazi, in order to assess the level of risk for an impending ventricular arrhythmia” (id.). The Examiner acknowledges “Armoundas fails to explicitly disclose the dynamically determined threshold value is adaptively changed based on a noise level in the detected cardiac electrical signals” (id.). 4 Appeal 2016-006425 Application 13/509,390 The Examiner finds Krishnamachari teaches a similar method for detecting altemans which incorporates a dynamic threshold dependent on random noise with a noise threshold. Further, it is disclosed that the voltage threshold is 1.8 microvolts and the noise threshold is 1.9 microvolts with application of a scale factor to the altemans noise value for a more effective voltage threshold value. (Ans. 3^4; emphasis omitted). The Examiner finds it obvious “to modify Armoundas to explicitly include a scaling factor incorporating compensation for a noise threshold, as taught by Krishnamachari, in order to effectively set the dynamic voltage threshold for alteman detection by adaptively taking into account noise in the cardiac electrical signals” {id. at 4; emphasis omitted). The issue with respect to this rejection is: Does the evidence of record support the Examiner’s conclusion that the prior art suggests “comparing the at least one repolarization altemans parameter to a dynamically determined threshold value that is adaptively changed based on a noise level in the detected cardiac electrical signals” as required by claim 1? Findings of Fact 1. The Specification teaches: Because the amplitude of random noise has a significant effect on the altemans voltage, noise, and K-score estimation, the altemans voltage, noise, and K-score threshold values are considered as functions of random noise levels. This allows for a more accurate assessment of repolarization altemans in the presence of significant random noise. (Spec. 1 88). 2. The Specification teaches an example where 5 Appeal 2016-006425 Application 13/509,390 a new altemans voltage threshold can be formed as a function of altemans noise for any value of random noise using an adapted version of Eqn. (3). For example, for random noise values of five microvolts, and for altemans noise of 1.8 microvolts, the corresponding altemans voltage threshold is 20.29 microvolts. (Spec. 177). 3. Armoundas teaches: In a method in accordance with the present invention, the level of repolarization altemans can be quantified by means well known in the art, such as measurement of the altemans voltage and measurement of the altemans ratio in one or more electrocardiographic leads. Threshold values of these parameters can be established such as 1.9 microvolts for the altemans voltage and a value of 3.0 for the altemans ratio. When the level of repolarization altemans exceeds a threshold value over some period of time (such as one minute) therapy is delivered to suppress the repolarization altemans and thus reduce the likelihood that a heart rhythm disturbance will occur. Repolarization altemans can be reliably estimated by analysis of approximately 128 beats. Thus, in about a minute or so (assuming a rate of 105-110 beats/min) the number of beats needed in the estimation will have been detected and/or recorded. As previously defined, repolarization altemans as used herein includes any change in the morphology of the T- wave or ST segment of the electrocardiogram on occurring on an every other beat basis. In other embodiments, the beat-to- beat variability in the cardiac electrical activity that is measured is heart rate variability. (Armoundas 139). 4. Armoundas teaches “a threshold value of heart variability may be established, such as the Standard Deviation of Normal to Normal RR intervals measure of heart rate variability being equal to 60 milliseconds” (Armoundas 139). 6 Appeal 2016-006425 Application 13/509,390 5. Farazi teaches: “It is also possible to have multiple thresholds such that in addition to determining whether T-wave altemans are present, changes in magnitudes of alternations can be determined. This can be used, e.g., to determine a degree of the T-wave altemans” (Farazi 13:13—17). 6. Farazi teaches: at least one metric is measured of T-waves in a predetermined number of beats (e.g., 2 to 10 beats) that follow each of the detected intrinsic premature contractions of the ventricles. The metric measured at step 304 can be, e.g., T-wave amplitude, T- wave slope, T-wave area, T-wave width, T-wave timing, T- wave morphology, QT interval, evoked QT interval, T-wave positive/negative slopes and durations, etc. (Farazi 11:40-47). 7. Christini teaches: If the sign is not consistent, as tested at step 160, then the amplitude of additional beats in a series of heart beats is again detected, as indicated by the arrow looping back to step 110. The heart beat processing is then repeated to dynamically define an electrical stimulus or stimuli to selectively apply through one or more electrodes. Only if the sign is consistent, however, is the electrical stimulus/stimuli delivered to the patient. (Christini 6:53—60). 8. Krishnamachari teaches: The ECG system 1020 may obtain thirty-two signals from the fourteen leads . . . These thirty-two signals may be used by the ECG system 1020 to produce a set of low-noise ECG signals (e.g., VI, V2, V3, V4, V5, and V6). The ECG system 1020 then may use the set of low-noise ECG signals to produce low noise vector signals (VM, X, Y, Z) using, for example, a vector enhancement technique. In some instances, the ECG system 1020 may produce other combinations of low- noise ECG signals. The ECG system 1020 also may associate a 7 Appeal 2016-006425 Application 13/509,390 high noise flag with the ECG signals. The ECG System 1020 may set the high noise flag if noise in the vector magnitude ECG lead VM is greater than HlGH_NOfSE_THRESHOLD, a value of about 1.8 pV, and if the noise level plus the altemans level in VM is greater than NOfSE_ALT_THRESHOLD, a value of about 2.5 pV. The ECG system 1020 then may generate altemans measures 1005 based on the low-noise ECG signals. For example, the ECG system 1020 may generate altemans measures 1005 through power spectmm (Fourier) analysis of the signals. Alternatively, time domain analysis may be employed, such as, for example, complex demodulation, estimation by subtraction, least squares estimation, autoregressive estimation, and/or auto-regressive moving average estimation. (Krishnamachari 5:30—54). Principles of Law “An examiner bears the initial burden of presenting a prima facie case of obviousness.” In re Huai-Hung Kao, 639 F.3d 1057, 1066 (Fed. Cir. 2011). A proper § 103 analysis requires “a searching comparison of the claimed invention — including all its limitations — with the teaching of the prior art.” In re Ochiai, 71 F.3d 1565, 1572 (Fed. Cir. 1995). Analysis Appellants contend “the combination of Armoundas, Farazi, Christini, and Krishnamachari fails to teach or suggest the dynamic thresholding recited in claim 1” (App. Br. 3). In particular, Appellants explain “Armoundas simply uses the level of RA as a binary trigger for whether or not to deliver therapy”; “Farazi is limited to teaching that multiple, static thresholds can be used to assess changes in T-wave altemans”; “Christini reveals that no mention is made of a threshold value that is changed as a 8 Appeal 2016-006425 Application 13/509,390 function of noise present in a measured signal”; and “[t]he portion of Krishnamachari cited in the Office Action (i.e., column 5, lines 30—54) does not teach adaptively changing a threshold for a repolarization altemans parameter based on the noise level in detected cardiac electrical signals” (App. Br. 3—5). The Examiner responds: Armoundas does disclose at least one repolarization altemans parameter (i.e., altemans ratio) that is adaptively changed based on a noise level in the detected cardiac electrical signal. The altemans ratio is well known in the art to be the magnitude of the alteman(s) divided by the noise level for at the time of the alteman (e.g., the noise level for that segment of data containing the alteman or the noise level determined by the current segment and all segments preceding the current segment). (Ans. 9). The Examiner finds “Krishnamachari further discloses the evaluation system 1030 modulating a determination of an alteman threshold crossing (and subsequent determination of sustained repolarization altemans) according to real-time ECG noise levels and/or noise levels relative to altemans levels” (id.). The Examiner also finds “Farazi discloses the well-known process of detecting intrinsic premature contractions of the ventricles by T-wave metrics to determine the presence of T-wave altemans. Thresholds are set due to actual changes in a metric of T- waves accounting for changes due to noise and/or changes due to misalignments” (id. at 10). We find that Appellants have the better position. Armoundas, as the Examiner conceded in the rejection (see Ans. 3), never suggests dynamic threshold values based on noise levels. The Examiner, in the response, contends the “altemans ratio is well known in the art to be the magnitude of 9 Appeal 2016-006425 Application 13/509,390 the alteman(s) divided by the noise level for at the time of the alteman” (Ans. 9), and newly cites Walker8 to support this position. We view this taking of Official Notice in the Answer, and the new reliance on Walker in the Answer,9 as improper. See In re Hoch, 428 F.2d 1341, 1342 n.3 (CCPA 1970) (“Where a reference is relied on to support a rejection, whether or not in a ‘minor capacity,’ there would appear to be no excuse for not positively including the reference in the statement of the rejection.”) Moreover, the Examiner has not established or identified a teaching in Walker demonstrating that this ratio serves as a dynamic threshold as required by claim 1. We also do not find a suggestion of dynamic threshold values based on noise levels in Farazi, Christini, or Krishnamachari in the portions relied upon by the Examiner (FF 5—8). In particular, the Examiner relies upon Krishnamachari for “application of a scale factor to the altemans noise value for a more effective voltage threshold value” (Ans. 4; emphasis omitted). However, as Appellants point out (see App. Br. 6), Krishnamachari sets a noise flag based on two specific threshold values, 1.8 pV and 2.5 pV, and does not dynamically adjust these thresholds based on the noise levels (FF 8). 8 Walker et al., Repolarization altemans: implications for the mechanism and prevention of sudden cardiac death, 57 Cardiovascular Res. 599-614 (2003). 9 See MPEP § 1207.03 (III) (“If Evidence (such as a new prior art reference, but not including a newly relied upon dictionary definition) is applied or cited for the first time in an examiner’s answer, then 37 CFR 41.39(a)(2) requires that the rejection be designated as a new ground of rejection.”) 10 Appeal 2016-006425 Application 13/509,390 Conclusion of Law The evidence of record does not support the Examiner’s conclusion that the prior art suggests “comparing the at least one repolarization altemans parameter to a dynamically determined threshold value that is adaptively changed based on a noise level in the detected cardiac electrical signals” as required by claim 1. B. 35 U.S.C. § 103(a) over Christini and Zhu The Examiner finds “Christini discloses a method for preemptively suppressing a heart rhythm disturbance in a patient’s heart with an implanted device configured to generate electrical impulses” but “Christini fails to explicitly disclose identifying the optimal location for detecting altemans” (Ans. 5—6). The Examiner finds: Zhu discloses detecting IEGM signals from multiple electrodes within the heart and choosing the optimal electrodes for pacing and detection of the capture based on signals from electrodes providing the highest probability of success for capture (Abstract; Col. 7, Line 48 - Col. 8, Line 57). To elaborate, Zhu explicitly discloses detection of loss of capture/detected capture, e.g., potentially due to lead dislodgement or exit block, leading to a switch to other sensing/pacing channels found to not employ that lead. Zhu is changing sensing/pacing configuration to one that best achieves capture AND sensing of said capture (i.e. by going down the list of preferred electrode configurations). (Ans. 6—7; emphasis omitted). The issue with respect to this rejection is: Does the evidence of record support the Examiner’s conclusion that the prior art suggests 11 Appeal 2016-006425 Application 13/509,390 “identifying optimal locations in a patient’s heart” for pacing electrodes as required by claim 21? Findings of Fact 9. Zhu teaches: “Clinical testing of the patient while the device is operating is commonly employed to select a particular pacing configuration and mode that is optimum for that particular patient” (Zhu 7:59-62). 10. Zhu teaches: “Capture verification can be performed by delivering a pacing pulse and attempting to sense an evoked response ... In this test, it is determined whether or not a sensing/pacing channel is achieving capture with the pacing pulses delivered by the channel’s electrode” (Zhu 6:53—63). 11. Zhu teaches: “Upon detection of a loss of capture in a sensing/pacing channel making up the selected pacing configuration, the controller is programmed to switch from the selected pacing configuration and associated pacing mode to a next pacing configuration and associated pacing mode contained in the ordered list” (Zhu 8:34-40). Analysis Appellants contend “Zhu does not teach or suggest the optimal placement of electrodes at locations where RA measurements can be maximized. Rather, Zhu is concerned with selecting the desired pacing mode for delivering treatment” (App. Br. 8). The Examiner responds Zhu teaches multiple electrodes located at a variety of pacing/sensing locations within the heart for pacing and sensing. Each electrode can be utilized for pacing or sensing and are connected to a switching circuit 70 (Fig. 1). The input of the evoked response sensing channel is preferably switched 12 Appeal 2016-006425 Application 13/509,390 to an electrode of another sensing/pacing channel for sensing (rather than utilizing the same electrode for sensing and pacing). Capture verification testing involves selecting (via the switching circuit 70) the particular electrode(s) utilized for evoked response detection in accordance with which electrode is placed in a location where an evoked response can be most easily sensed (Col. 6, Line 53 - Col. 7, Line 47). (Ans. 10). We find that Appellants have the better position. At no point in the Examiner’s response does the Examiner establish that Zhu or Christini teach “identifying optimal locations in a patient’s heart” followed by placing leads “at the identified optimal locations” as required by claim 21. Instead, Christini teaches placing leads in the heart (FF 7) and Zhu appears to teach optimizing the configuration of leads that have already been implanted (FF 9-11). To the extent that the Examiner relies upon Zhou10 as evidentiary, the Examiner does not identify a teaching in Zhou for either “identifying optimal locations in a patient’s heart” or for placing leads “at the identified optimal locations” in the Answer (see Ans. 10—11). Conclusion of Law The evidence of record does not support the Examiner’s conclusion that the prior art suggests “identifying optimal locations in a patient’s heart” for placing electrodes as required by claim 21. C. 35 U.S.C. § 103(a) over Christini, Zhu and Sharma This rejection relies upon the underlying obviousness rejection over Christini and Zhu. Having reversed the Christini and Zhu rejection for the 10 Zhou et al., US 2006/0116596 Al, published June 1, 2006. 13 Appeal 2016-006425 Application 13/509,390 reasons given above, we necessarily reverse the obviousness rejection further including Sharma, since the Examiner does not rely upon Sharma to address the “identifying optimal locations in a patient’s heart” limitation. SUMMARY In summary, we reverse the rejection of claims 1—5 and 10—20 under 35 U.S.C. § 103(a) as obvious over Armoundas, Farazi, Christini, and Krishnamachari. We reverse the rejection of claims 7 and 21 under 35 U.S.C. § 103(a) as obvious over Christini and Zhu. We reverse the rejection of claims 8 and 9 under 35 U.S.C. § 103(a) as obvious over Christini, Zhu, and Sharma. REVERSED 14 Copy with citationCopy as parenthetical citation