BOSTON SCIENTIFIC SCIMED, INC.Download PDFPatent Trials and Appeals BoardNov 19, 20212021002922 (P.T.A.B. Nov. 19, 2021) Copy Citation UNITED STATES PATENT AND TRADEMARK OFFICE UNITED STATES DEPARTMENT OF COMMERCE United States Patent and Trademark Office Address: COMMISSIONER FOR PATENTS P.O. Box 1450 Alexandria, Virginia 22313-1450 www.uspto.gov APPLICATION NO. FILING DATE FIRST NAMED INVENTOR ATTORNEY DOCKET NO. CONFIRMATION NO. 15/956,824 04/19/2018 Nicholas Lee Tassoni 2001.1780101 2321 11050 7590 11/19/2021 SEAGER, TUFTE & WICKHEM, LLP 100 South 5th Street Suite 600 Minneapolis, MN 55402 EXAMINER BYRD, BRIGID K ART UNIT PAPER NUMBER 3771 NOTIFICATION DATE DELIVERY MODE 11/19/2021 ELECTRONIC Please find below and/or attached an Office communication concerning this application or proceeding. The time period for reply, if any, is set in the attached communication. Notice of the Office communication was sent electronically on above-indicated "Notification Date" to the following e-mail address(es): BSC.USPTO@stwiplaw.com PTOL-90A (Rev. 04/07) UNITED STATES PATENT AND TRADEMARK OFFICE ____________ BEFORE THE PATENT TRIAL AND APPEAL BOARD ____________ Ex parte NICHOLAS LEE TASSONI, KHOI LE, and JEFFRY D. JOHNSON Appeal 2021-002922 Application 15/956,824 Technology Center 3700 ____________ Before JOHN C. KERINS, MICHAEL L. HOELTER, and ANNETTE R. REIMERS, Administrative Patent Judges. HOELTER, Administrative Patent Judge. DECISION ON APPEAL STATEMENT OF THE CASE Pursuant to 35 U.S.C. § 134(a), Appellant1 appeals from the Examiner’s decision to reject claims 1–18 and 20. See Final Act. 1 (Office Action Summary). We have jurisdiction under 35 U.S.C. § 6(b). We AFFIRM the Examiner’s rejections of these claims. 1 We use the word “Appellant” to refer to “applicant” as defined in 37 C.F.R. § 1.42. Appellant identifies the real party in interest as “Boston Scientific SciMed, Inc.” Appeal Br. 3. Appeal 2021-002922 Application 15/956,824 2 CLAIMED SUBJECT MATTER The disclosed subject matter “pertains to medical devices and methods for manufacturing and/or using medical devices.” Spec. 1:10–11. System claims 1 and 17, and method claim 20, are independent. Claim 1 is illustrative of the claims on appeal and is reproduced below. 1. A vascular occlusion system, comprising: a microcatheter configured to navigate a vasculature; an elongate shaft slidably disposed within a lumen of the microcatheter, the elongate shaft having a lumen extending from a proximal end of the elongate shaft to a distal end of the elongate shaft; a plurality of occlusive medical devices releasably connected to the elongate shaft in series; and a release wire slidably disposed within the elongate shaft and at least a portion of each of the plurality of occlusive medical devices; wherein the release wire sequentially secures the plurality of occlusive medical devices to the distal end of the elongate shaft when the release wire is disposed within at least a portion of each of the plurality of occlusive medical devices, wherein the release wire includes one or more indicators proximate a proximal end of the release wire configured to communicate to a user of the vascular occlusion system how many of the plurality of occlusive medical devices have been released when the release wire is partially withdrawn from the elongate shaft. EVIDENCE Name Reference Date Buiser et al. (“Buiser”) US 2007/0083226 A1 Apr. 12, 2007 Lee at al. (“Lee”) US 2014/0236127 A1 Aug. 21, 2014 Jones et al. (“Jones”) US 2015/0335333 A1 Nov. 26, 2015 Appeal 2021-002922 Application 15/956,824 3 REJECTIONS Claims 1–13, 17, 18, and 20 are rejected under 35 U.S.C. § 103 as unpatentable over Buiser and Lee. Claims 14–16 are rejected under 35 U.S.C. § 103 as unpatentable over Buiser, Lee, and Jones. ANALYSIS The rejection of claims 1–13, 17, 18, and 20 as unpatentable over Buiser and Lee Appellant argues the three independent claims (i.e., claims 1, 17, and 20) together. See Appeal Br. 8–11. Appellant’s arguments with respect to dependent claims 2–13 and 18 are that they depend from an allowable independent claim. See Appeal Br. 11. We select claim 1 for review, with the remaining claims (i.e., claims 2–13, 17, 18, and 20) standing or falling with claim 1. See 37 C.F.R. § 41.37(c)(1)(iv) (2019). The Examiner primarily relies on the teachings of Buiser for disclosing the limitations of claim 1, including the disclosure of “a plurality of occlusive medical devices” and “a release wire (control wire 106).” Final Act. 5 (referencing Buiser ¶¶ 47, 62). However, the Examiner acknowledges that “Buiser fails to explicitly disclose wherein the release wire includes one or more indicators” used to communicate “how many of the plurality of occlusive medical devices have been released” as recited. Final Act. 7. The Examiner relies on Lee for also teaching a release wire but importantly, “wherein the release wire includes one or more indicators” as recited. Final Act. 7 (referencing Lee ¶ 48). The Examiner, thereafter, provides a reason for combining Buiser and Lee, i.e., it would have been obvious “to modify Buiser’s device to include indicators, as taught by Lee, Appeal 2021-002922 Application 15/956,824 4 in order to alert the operator when the plurality of occlusive medical device have been released from the delivery device,” and also “how many.” Final Act. 7. Appellant disagrees contending that Lee fails to provide “indication of how many of [the] medical devises have been released, because the reference teaches a system with a single implant.” Appeal Br. 9. In other words, “Lee fails to teach or suggest their system as being capable of being used to deliver multiple implants” because “Lee teaches a single implant.” Appeal Br. 9. Appellant’s contentions are not persuasive of Examiner error because the Examiner relies on Buiser for disclosing the release of multiple implants, not Lee.2 See, e.g., Buiser ¶¶ 47 (“first embolic coil 16” and “second embolic coil 26”), 64 (using a control wire for “pushing the coils out of the catheter or retracting the catheter”). The Examiner relies on Lee as teaching providing an indicator on a control wire to signal that a medical device has been released, and determines that the person of ordinary skill in the art would recognize that multiple indicators could be used (“capable of communicating . . . how many”) if multiple medical devices are to be released. Final Act. 7; see also Lee ¶ 48 (discussing disengaging “the male coupler 56 from the female coupler 40” and that “the control member 70 may include a marker (not shown) near its proximal end to alert the operator when the male coupler 56 has been completely withdrawn from the female coupler 40”). Thus, Appellant’s focus on Lee for purportedly pertaining to a 2 Appellant acknowledges, “Buiser teaches a system involving multiple coils.” Appeal Br. 9. Appeal 2021-002922 Application 15/956,824 5 single implant is not commensurate with the Examiner’s rejection, or the teachings the Examiner relied on Lee for. Appellant also contends the Examiner relied on “hindsight” in reaching this rejection. See Appeal Br. 9; see also Reply Br. 3 (“the only rational reason for modifying Buiser with Lee is based on impermissible hindsight”). We disagree because, as expressed by the Examiner, Lee’s use of a marker to indicate when a male/female uncoupling occurred would “teach an indication of how many . . . devices have been released when [Buiser’s] release wire is partially withdrawn.” Final Act. 7–8 (referencing Buiser ¶¶ 62, 64). Appellant does not explain how using a marker on a wire to designate a release as taught by Lee was gleaned only from Appellant’s Specification. Appellant further argues that there is no reason to combine Buiser and Lee because “Buiser already teaches using radiopaque markers on the coils,” and as such, “[t]he addition of the marker of Lee would appear to be redundant to the system of Buiser.” Appeal Br. 10; see also Buiser ¶ 77 (“[a]t least one of the coils may include a radiopaque marker”). In other words, as per Appellant, “Buiser already teaches markers that would communicate to the user how many of the plurality of coils have been released.” Appeal Br. 11. The Examiner responds, the engaging members of the coils cannot be retracted or moved until the control wire extending through the engaging members is retracted (Para. [0062] of Buiser). Therefore, one of ordinary skill would’ve understood indication means comprised specifically on the control wire to be useful, since the control wire has to be retracted a sufficient distance to deploy the coils. Ans. 4. Continuing in this vein, the Examiner reasons Appeal 2021-002922 Application 15/956,824 6 Therefore, one of ordinary skill would’ve understood the indication means of Lee could be included on a control wire, in a specific location, to communicate when an implant, or multiple implants, have been deployed due to retraction of the control wire, which would communicate how many implants would be deployed. Ans. 4. The Examiner also states, “the motivation is to alert the operator when the implant has been uncoupled from the delivery device.” Ans. 4. Appellant disagrees with this assessment stating that Buiser’s radiopaque “markers could be viewed during deployment to determine the position of each marked coil/engaging member relative to the aneurysm.” Reply Br. 2. Appellant continues stating “[h]aving the markers on the coils/engaging members would provide the operator the ability to see where each marked coil was positioned.” Reply Br. 2. We agree with Appellant that the radiopaque marker would enable a user to ascertain the position of the medical device when viewed on a screen, but the Examiner’s rationale is not premised on the position of the medical device, but on whether or not the medical device has been deployed or uncoupled from the installing catheter. Lee specifically teaches the use of a marker for this exact purpose, i.e., the ability to ascertain when an uncoupling has occurred (see Lee ¶ 48), regardless of where in the body the device may be located. We thus do not fault the Examiner for finding that a skilled person “would’ve understood indication means comprised specifically on the control wire to be useful,” such as “to communicate when an implant, or multiple implants, have been deployed.” Ans. 4. Appellant’s further argument contending that by placing markers on the control wire, the Examiner “would remove this feature” (i.e., remove Buiser’s radiopaque coating) (Reply Br. 2) is not only unsupported by the Appeal 2021-002922 Application 15/956,824 7 record but is also without merit because there is no reason they cannot each be employed for their own special purpose. Appellant also argues that “the single marker would still fail to provide any information regarding how many implants were released” (Reply Br. 2), but again, Appellant is too narrowly applying Lee’s teaching as something of a bodily incorporation in which only a single marker is ever employed, and does not account for the Examiner’s use of Lee’s disclosure of using a marker to indicate release of a medical device, as applied to the release of a plurality of devices, and using a corresponding plurality of markers to indicate the separation of each of the devices. In summation, and based on the record presented, we sustain the Examiner’s rejection of claims 1–13, 17, 18, and 20 as being obvious in view of Buiser and Lee. The rejection of claims 14–16 as unpatentable over Buiser, Lee, and Jones Appellant argues dependent claims 14–16 together. See Appeal Br. 11–12; Reply Br. 3. We select claim 14 for review, with claims 15 and 16 standing or falling therewith. Claim 14 depends directly from claim 1 and further recites “a second release wire” disposed within the catheter. The Examiner relies on the additional reference to Jones for teaching “a second release wire (engagement member 170)” as recited. Final Act. 21. The Examiner reasons that it would have been obvious to combine Jones’ teachings with that of Buiser and Lee “in order to maintain secure coupling of the embolic coil with the delivery member until desired release of the coil, which would prevent unwanted release of the coils until reaching the target location Appeal 2021-002922 Application 15/956,824 8 within the body.” Final Act. 21; see also Ans. 4 (both referencing Jones ¶ 76). Appellant disputes that there is a “rational reason” for combining Jones with that of Buiser and Lee stating that adding Jones’ second release wire for security purposes to prevent unwanted release is improper “as Buiser already teaches a structure that achieves these goals.” Appeal Br. 11, 12. Appellant contends that “Buiser already teaches the coil(s) being securely coupled to the pusher wire 46 via the engaging members 22 (embodiment in FIGS 1 and 2A) and also teaches the coil 128 being securely coupled to the pusher wire 130 via the engaging members 132 and the control wire 106 (embodiment in FIG 7A).” Appeal Br. 12. In each of these embodiments, Buiser maintains inter-coupling via the sole wire extending through the corresponding engaging members, the same as depicted in Appellant’s Figures 2–5. This single wire is the only bulwark against an undesired deployment depicted in Buiser. However, the Examiner is relying on Jones for its teaching of “a second release wire” to “prevent unwanted release” by inadvertent withdrawal of Buiser’s sole wire “until reaching the target location within the body.” Final Act. 21 (referencing Jones Fig. 3B, item 170, and ¶ 76); see also Ans. 4–5. Appellant does not explain why the desire to insure against unwanted deployment is not a “rational reason” to do so. Appeal Br. 11. We further note that Jones’ teaching of additional security is quite similar to, structurally and operationally, Appellant’s solution. Compare Jones Fig. 3B and ¶ 76, with Appellant’s Figs. 6, 7 and Spec. 22:27 to 24:24. Hence, Appellant’s overly optimistic characterization of Buiser’s couplers above, while at the same time discussing a manner of achieving greater security Appeal 2021-002922 Application 15/956,824 9 with respect to Appellant’s couplers, is not persuasive that the Examiner erred by seeking to restrain Buiser’s couplers with the same level of security taught by Jones. See Ans. 5 (“Buiser could be modified with Jones to prevent unwanted release of the coils” such as by “preventing unwanted retraction” of Buiser’s control wire). In Appellant’s Reply Brief, Appellant takes a different approach, focusing upon the embodiment depicted in Figure 8 of Buiser. See Reply Br. 3. However, there is no indication why this focus on this different embodiment could not have also been presented in Appellant’s Appeal Brief when Appellant discussed the embodiments of Figures 1, 2A, and 7A of Buiser (as noted above and which were addressed by the Examiner). On this point, 37 C.F.R. § 41.41(b)(2) states, “[a]ny argument raised in the reply brief which was not raised in the appeal brief . . . will not be considered by the Board for purposes of the present appeal, unless good cause is shown.” Here, good cause has not been shown as to why this Figure 8 embodiment of Buiser could not have also been addressed when Appellant discussed the Figure 1, 2A, and 7A embodiments of Buiser. We thus decline to consider this new argument focusing on this different embodiment, especially when the Examiner was not provided an opportunity to respond to this different embodiment when responding to the other embodiments. Appellant also contends that there is “no rational reason” for the combination with Jones “other than looking to Appellant’s specification, which is an error.” Appeal Br. 11. However, we do not abide by Appellant’s charge of “hindsight” when the feature identified and the benefits achieved are already known in the art. Appeal 2021-002922 Application 15/956,824 10 Appellant also contends that “Jones fails to teach any configuration in which either member is prevented from being proximally retracted.” Reply Br. 3. However, this is not a correct reading of cited paragraph 76 of Jones (see Final Act. 20, 21; Ans. 4), which specifically teaches that Jones’ corresponding “second release wire” 170 (Final Act. 21) is “slidably positioned” within the delivery member. Note also Jones’ discussion of other “release member 180” being retracted from a “first configuration” (where its distal end is in engagement with the distal end of wire 170 (see Jones Fig. 3)) to a “second configuration” where member 180 “is moved proximally.” Jones ¶ 76. Hence, Appellant’s contention regarding the lack of movement of Jones’ members is not persuasive of Examiner error. Further, and of perhaps greater importance, is the fact that claim 14 is devoid of language requiring either member to be “prevented from being proximally retracted” as argued by Appellant. Accordingly, and based on the record presented, we sustain the Examiner’s rejection of claims 14–16 as being obvious in view of Buiser, Lee, and Jones. CONCLUSION In summary: Claim(s) Rejected 35 U.S.C. § References/Basis Affirmed Reversed 1–13, 17, 18, 20 103 Buiser, Lee 1–13, 17, 18, 20 14–16 103 Buiser, Lee, Jones 14–16 Overall Outcome 1–18, 20 TIME PERIOD FOR RESPONSE Appeal 2021-002922 Application 15/956,824 11 No time period for taking any subsequent action in connection with this appeal may be extended under 37 C.F.R. § 1.136(a)(1)(iv). AFFIRMED Copy with citationCopy as parenthetical citation